top of page

Beamline 7.2W
Macromolecular Crystallography
Personnel Safety Interlock (PSI) System
🔒 Closing the Hutch Door


-
Press the orange "Search" button
-
Beacon flashes orange & green
-
The alarm sounds for 15 seconds
2. Ensure all personnel exit the hutch

1
2
3. Close and lock the hutch door
Turn key anticlockwise to 12 o’clock
4. Insert key into door lock slot and turn to ON
-
Must complete within 15 seconds
-
Beacon turns orange if successful
-
If timeout: beacon turns green → restart from step 1

1

2

5. Press green "ON" button on magnetic door lock

1

2
6. Push door firmly closed
-
Green light = magnetic lock engaged
-
Red light = lock failure → PSI blocks shutter


7. Turn on MBS
-
Confirm "X-RAY ON" light is lit
-
Beacon turns red
Option 1

Option 2

bottom of page